To ensure a longer service life for an X-ray baggage scanner, proper maintenance is crucial. Regular maintenance can prevent major issues and ensure the machine operates smoothly. Here are some essential daily maintenance tips and techniques for X-ray baggage scanners:
1. Keep Environment Dry
X-ray baggage scanner are made of electronic products, circuit boards and other materials. The main components are organic electricity, electronics, and optics. Wait. If after being damp due to the influence of the humid environment, the lighter will cause circuit parameter changes or mechanical parts movement failure, and the more serious will cause the electrical components to become mildewed and short-circuited after being damp, which will burn out.
We must take moisture-proof measures to keep the environment dry and prevent the X-ray baggage scanner from being damp and mildewed. In this way, the normal operation of the machine can be ensured, and the service life of the machine can be prolonged. Moisture prevention is one of the safety measures and must be taken seriously.
2. External Cleaning of X-ray Baggage Scanner
Because the X-ray baggage scanner is used to detect passengers’ luggage, it will inevitably be dusty if the machine is placed in a public place for a long time. You should also pay attention to the X-ray baggage scanner when scrubbing it. If you use a wet towel to scrub the equipment, You don’t need water or less water to measure, and you don’t need to wet the cloth to scrub the machine. Just wipe off the dust with a dry towel. If the X-ray baggage scanner is too dirty, the towel should not be too wet when scrubbing, wring out the towel before wiping. Otherwise, the water from the towel will flow into the X-ray baggage scanner to make the machine damp.
3. Internal Cleaning of X-ray Baggage Scanner
The X-ray baggage scanner must not only be moisture-proof but also dust-proof. If dust invades the inside of the machine, it will also cause a short circuit during the shutdown of the security inspection X. Therefore, dust-proofing is also an important measure to ensure the normal operation of the machine. When cleaning the external dust, it is best to use a vacuum cleaner; while the dust inside the machine, it is best to use a hairdryer and a fine brush to clean up, never wipe it with a damp cloth. Some parts can be covered with a cloth cover to achieve a very good dustproof effect.
4. Operate the X-ray Baggage Scanner
The X-ray Baggage Scanner with omnidirectional wheels, allowing it to move smoothly on flat surfaces without causing damage. However, please be careful to avoid severe collisions, as these may cause misalignment of the X-ray Radiation Source and damage the hardware.
5. Pay Attention to Safety Check
Make sure the electrical connections are secure. Make sure there are no damaged wiring or rats living inside. Make sure the emergency stop button is not pressed before operation. If the power supply in the area is unstable and there will be intermittent power outages, a UPS is required.
Normal maintenance work
1. The equipment should be installed in a ventilated, dusty, dry environment, avoid high temperature, humidity and direct sunlight.
2. During maintenance, remove the dust from the equipment first, and remove the dust with a brush and a vacuum cleaner.
3. If there is any looseness in the component or the terminal, it must be tightened immediately.
4. When the components are corroded, replace them if necessary.
5. If any component is damaged, find out the cause, and replace the new component after troubleshooting. After the device is replaced, the relevant parameters must be adjusted before it can be used normally.
6. Do not allow foreign objects or liquids to enter the equipment during maintenance to prevent leakage or accidents.
7. The specific content of maintenance and maintenance
8. Check the power indicator on the top of the device: When the device is powered on, the green indicator light is on.
9. Check the X-ray indicator light on the top of the equipment: when the ray is emitted, the red indicator light is on.
10. Check the emergency stop button above the channel: press the emergency stop button, the equipment can be powered off immediately. When the button is not reset, the device cannot be restarted.
11. Check the key switch above the channel: After the key switch is turned on, the start button can start the equipment. The radiation stop button can disconnect the power of the radiation controller. If the key switch becomes less sensitive, replace it